Foodstuffs
The food category encompasses a wide range of substances with varying properties.
It includes agricultural, livestock and marine products and processed products, as well as food additives. The moisture content of these items also varies widely. It is necessary, therefore, to select the most appropriate sample handling procedures and measurement method for each individual product.
Test methods for foodstuffs are mostly stipulated in the JAS standards. In most cases moisture content is measured using heating and drying methods, but Karl Fischer titration is specified for some items. In the case of chemically defined foodstuffs, such as sugars and additives, Karl Fischer titration can be applied using the same methods as for normal chemicals.
Food items with complex structures, such as cereals and meats, contain not only adhesion moisture but also moisture incorporated within cells.
Because many food items do not dissolve in organic solvents, the moisture vaporization method is used extensively. However, foodstuffs commonly become unstable when heated, and water may be formed as substances break down. Care must be taken, therefore, when setting analytical conditions.
Food Additive and Flavorings
Cereals and Dried Vegetables
Sugars and Condiments
Confectionery
Diary Products
Gourmet Foods
Food Additives and Flavorings Key Points
Food additives are substances used in the manufacture, processing and preservation of foodstuffs. There are many types with varying attributes. Most are solid, but those that dissolve in dehydrated solvents or electrolytes can be titrated directly using Karl Fischer titration.
First, it is necessary to determine whether or not interference substances are present.
Direct titration should be avoided in the case of insoluble substances; the moisture vaporization method is recommended.
As the moisture content of these substances is generally high, volumetric titration is used.
Coulometric titration can also be used if the sample quantity is kept small.
When using moisture vaporization, it is necessary to set the measurement conditions after determination of the heating temperature. Official specifications for food additives commonly stipulate the loss-on-drying method, so conformance with that method should be checked.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, SU (or MS, FM) 25-50ml
|
|
Substance
|
Dehydrated
Solvents
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(ppm)
|
(Flavorings) Orange Powder
|
General-use
|
0.0228
|
10.96
|
48.0
|
(Condiments) Sodium Citrate
|
General-use
|
0.1711
|
20.97
|
12.3
|
(Emulsifiers) Sorbitan Fatty Acid Ester
|
General-use
|
2.0962
|
2.95
|
0.141
|
(Sweeteners) Aspartame
|
Sugar-type
|
0.0535
|
1.33
|
2.49
|
(Preservatives) Sorbic Acid
|
General-use
|
1.0059
|
1.33
|
2.49
|
|
(Strengtheners) Calcium Lactate
|
Sugar-type
|
0.0475
|
13.75
|
28.9
|
|
(Acidulants) Citric Acid
|
General-use
|
0.1044
|
9.23
|
8.84
|
|
(Water-retaining emulsifying stabilizer) Sodium Chondroitin Sulfate
|
Sugar-type
|
0.0483
|
4.49
|
9.30
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Aquamicron SS-Z (or SS) 150ml, Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100 ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
(Flavoriings) Vanilla Powder
|
130
|
0.1161
|
4.70
|
4.05
|
(Flavorings) Orange Powder
|
150
|
0.2943
|
6.01
|
2.04
|
|
(Strengtheners) Calcium Lactate
|
200
|
0.0513
|
14.50
|
28.3
|
|
|
|
|
Coulometric Titration
Reagents used: Aquamicron AX (or AS) 100ml, Aquamicron CXU 5ml
|
|
Substance
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(ppm)
|
Sodium Citrate
|
0.0785
|
9649
|
12.3
|
Saccharin
|
0.0921
|
12813
|
13.9
|
|
|
|
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content (%)
|
(Condiments) Sodium glutamate
|
180
|
0.0484
|
4739
|
9.79
|
(Sweeteners) Asparatame
|
150
|
0.0426
|
1001
|
2.35
|
|
(Preservatives) Sorbic Acid
|
160
|
1.0252
|
3495
|
0.341
|
|
(Strengtheners) Calcium Citrate
|
150
|
0.0940
|
12257
|
13.0
|
|
(Adiculants) Malic Acid
|
105
|
1.0389
|
2132
|
0.205
|
|
(Quality enhancers) Sodium Polyphosphate
|
130
|
2.0194
|
4988
|
0.247
|
|
(Water-retaining emylsifying stabilizer) Sodium Chondroitin Sulfate
|
150
|
0.0440
|
4182
|
9.50
|
|
| back to the top |
|
|
Cereals and Dried Vegetables Key Points
Volumetric titration is suitable for cereals and starches which have high moisture content (approximately 10%). If the sample is in powder form it can be dispersed in dehydrated solvent, allowing the moisture content to be extracted for direct Karl Fischer titration. When using a dehydrated solvent with sugars, solubility is sometimes enhanced if the temperature is raised to 40-45°C before titration. The moisture vaporization method is suitable for the many items that do not dissolve in solvents. Care needs to be taken with the heating temperature for moisture vaporization. Temperatures in the 130-180°C range are commonly used. High temperatures may cause various phenomena, including discoloration, abnormal results or failure to reach an endpoint. In such cases the measurement must be repeated at a lower temperature.
Moisture vaporization should be used with coulometric titration. Cereals cannot be titrated directly as they do not dissolve in electrolytes. Cereals are sometimes sampled after preliminary processing (e.g. pulverization). Care is needed to avoid moisture vaporization or absorption during these processes.
Volumetric titration is the principal method used for dried vegetables. The sample is added to the dehydrated solvent and stirred well for a period of 5-10 minutes before direct titration. Moisture vaporization is used when the moisture content cannot be fully extracted from samples in this way. Karl Fischer titration can be carried out using either the volumetric or coulometric methods. As moisture content tends to be high, volumetric titration is usually preferable.
The precautions stipulated for moisture vaporization in connection with cereals also apply to dried vegetables.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, SU (or MS, FM) 25-50ml
|
|
Substance
|
Dehydrated
Solvents
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(ppm)
|
Sweet Potato Starch
|
General-use
|
0.1885
|
31.10
|
16.5
|
Soy Protein
|
Sugar-type
|
0.0756
|
5.43
|
7.18
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Maize
|
130
|
0.0529
|
6.34
|
12.0
|
Apple Powder
|
110
|
0.1647
|
3.60
|
2.19
|
|
| back to the top |
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX (or AS) 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Maize
|
150
|
0.0482
|
7009
|
14.5
|
Soy Protein
|
150
|
0.0463
|
4876
|
8.10
|
|
|
|
|
|
Sugars and Condiments Key Points
Sugars are dissolved in a sugar-type dehydrated solvent and titrated. Volumetric titration is used. If a substance is slow or difficult to dissolve, the solvent can be heated (up to 50°C) and titrated. (If the temperature exceeds 50°C the formamide in the sugar-type dehydrated solvent will gradually break down. The Karl Fischer reagent will be consumed in small amounts leading to uncertainty about the end point.)
Condiments are normally dispersed in general-use dehydrated solvent. This is followed by extraction and titration. If the ingredients of a particular item give high sugar content, a sugar-type dehydrated solvent is used. An oil-type dehydrated solvent is used for products with a high fat content.
Moisture vaporization should generally be avoided for sugars as these tend to break down when heated. Moisture vaporization is unnecessary for condiments, since in almost all cases moisture content can be measured using direct titration.
When using the moisture vaporization method the optimal heating temperature must be determined first. Changes in the sample could lead to errors.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, SU OLII (or MS, FM) 25-50ml
|
|
Substance
|
Dehydrated
Solvents
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(ppm)
|
Roast Meat Gravy
|
General-use
|
0.0187
|
10.66
|
57.0
|
Glucose
|
Sugar-type
|
1.1052
|
1.64
|
0.148
|
Mayonnaise
|
Oil-type
|
0.1425
|
28.93
|
20.3
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Soy Sauce
|
200
|
0.0399
|
24.07
|
60.3
|
Mayonnaise
|
200
|
0.1408
|
28.61
|
20.3
|
|
|
|
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX (or AS) 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Sugar
|
130
|
0.4853
|
3669
|
1.98
|
Spices
|
130
|
0.0412
|
5830
|
41.2
|
|
|
|
|
|
Confectionery Key Points
Because sugar is the main ingredient of confectionery, these items are dissolved in a sugar-type dehydrated solvent for titration. Volumetric titration is used.
Coulometric titration is normally avoided as these products will not dissolve in the anolytes. The moisture vaporization method is also unsuitable as heating causes sugars to break down and produce water.
Caramels, candies and similar items are titrated after they have been dissolved in a sugar-type dehydrated solvent that has been heated (up to 50°C). (If the temperature exceeds 50°C the formamide in the sugar-type dehydrated solvent will gradually break down. The Karl Fischer reagent will be consumed in small amounts and this will lead to uncertainty about the end point.) The moisture vaporization method can be used for snack-type confectionery.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, SU (FM) at 40°C, 25-50ml
|
|
Substance
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(ppm)
|
Milk Caramel
|
0.2869
|
23.42
|
8.16
|
Chewing Gum
|
0.2978
|
5.63
|
1.89
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Potato Chips
|
130
|
0.1814
|
3.73
|
20.6
|
Corn Snacks
|
130
|
0.2462
|
3.44
|
1.40
|
|
|
|
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX (or AS) 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Small rice Crackers
|
130
|
0.0377
|
2776
|
7.36
|
|
| back to the top |
|
|
Dairy Products, Fat Products Key Points
Because of their high moisture content, most dairy products are titrated directly using the volumetric method. A sugar-type dehydrated solvent is normally used as the titration solvent. However, products that have fats as their principal ingredients, such as butter, are dissolved in an oil-type dehydrated solvent for titration. When volumetric titration is used for fat products they are dissolved in an oil-type dehydrated solvent.
The moisture content of vegetable oils, which contain little water, can be measured more accurately using coulometric titration.
If a product will not dissolve readily in the electrolyte, the addition of chloroform to the anolyte (equivalent to 20% by volume of anolyte) will allow the substance to dissolve for repeat titrations.
The moisture vaporization method can be used for samples of these types. However, the conditions must be determined with care. As a general rule, a heating temperature in the 110-180°C range is used.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, OLII, SU (or MS, CM, FM) 25-50ml
|
|
Substance
|
Dehydrated
Solvents
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Cream Cheese
|
General-use
|
0.0485
|
25.04
|
51.6
|
Fresh Cream
|
Sugar-type
|
0.1142
|
16.74
|
14.7
|
Butter
|
Oil-type
|
0.0303
|
4.25
|
14.0
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Cream Cheese
|
130
|
0.1147
|
59.14
|
51.06
|
Margarine
|
150
|
0.1281
|
19.97
|
15.6
|
|
|
|
|
Coulometric Titration
Reagents used: Aquamicron AX (or AS) 100ml, Aquamicron CXU 5ml
|
|
Substance
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Butter
|
0.0422
|
5845
|
13.9
|
Cheese
|
0.0142
|
5840
|
41.1
|
Soy Bean Oil
|
4.5118
|
29
|
51ppm
|
|
|
|
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX (or AS) 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Powdered Milk
|
110
|
0.0648
|
2953
|
4.56
|
Cheese
|
130
|
0.0217
|
8982
|
41.4
|
|
|
|
|
|
Gourmet Foods Key Points
Volumetric titration is normally used for gourmet foods because of its high moisture content. They are dissolved in a sugar-type dehydrated solvent and subjected to Karl Fischer titration. If a product will not dissolve readily the solvent can be heated to 40°C. An oil-type dehydrated solvent is used for high-fat foods. A general-purpose dehydrated solvent may be used for some samples. If the moisture content cannot be extracted fully using these solvents, Karl Fischer titration is used in conjunction with moisture vaporization. Either volumetric or coulometric titration can be used in such cases.
Direct coulometric titration should be avoided due to problems with sample solubility in the anolyte.
|
|
Examples of Measurement
Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX, OLII, SU (or MS, CM, FM) 25-50ml
|
|
Substance
|
Dehydrated
Solvents
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Coffee
|
Sugar-type
|
0.1287
|
3.87
|
3.01
|
soy Milk
|
Oil-type
|
0.0663
|
57.26
|
86.4
|
|
|
|
|
Moisture Vaporization - Volumetric Titration
Reagents used: Karl Fischer Reagent SS-Z (or SS), Dehydrated Solvent GEX (or MS) + PG (3:1) 50-100ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(mg)
|
Moisture
Content
(%)
|
Coffee Beans
|
130
|
0.1919
|
4.11
|
2.14
|
Dried Beer yeast
|
200
|
0.0840
|
8.86
|
10.6
|
|
|
|
|
Moisture Vaporization - Coulometric Titration
Reagents used: Aquamicron AX (or AS) 150ml, Aquamicron CXU 10ml
|
|
Substance
|
Heating
Temperature
(°C)
|
Sample
Quantity
(g)
|
Measurement
Value
(µg)
|
Moisture
Content
(%)
|
Coffee
|
90
|
0.1282
|
3229
|
2.52
|
Hops
|
130
|
0.0189
|
1536
|
8.14
|
Gelatin
|
180
|
0.0114
|
1790
|
15.7
|
|
|
|